ClearanceBrief
Recall Z-1811-2022 ยท PQF Sensor, Glucose, Invasive, Non-Adjunctive

Tandem Diabetes Care Inc

t:slim X2 insulin pump; t:slim X2 insulin pump with Dexcom G5; t:slim X2 insulin pump with Basal-IQ technology; t: slim x2 insulin pump with the Control-IQ Technology

DateSep 26, 2022 (initiated May 24, 2022)
StatusOpen, Classified
ReasonInsulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.
Root causeDevice Design
ActionOn 5/24/22, and 7/7/22, correction notices sent to customers in Australia and New Zealand to update to the latest software. t:slim X2 insulin pump with Basal-IQ technology version 6.6 and Control-IQ technology version 7.6 available with latest software versions. Software update contact: Australia, AMSL: diabetes@amsl.com.au or 1300 851 056; New Zealand, NZMS tandemupdate@nzms.co.nz (Questions: 0508 634 103) On 7/26/22, correction notices were mailed or emailed in the United States to healthcare providers and distributors. Recalling firm contact: Techsupport@tandemdiabetes.com Providers asked
Quantity31,100
DistributionUS: NC, OK, ME, NY, TX, LA, NE, FL, AZ, IL, OH, CA, AL, MI, UT, VA, MA, SC, MS, GA, NV, HI, MN, WA, IN, AR, ID, CO, IA, MO, MT, PA, WI, NH, MD, WV, KS, NM, OR, KY, DE, ND, CT, NJ, TN, WY, AK, SD, DC, RI, IA, PR, VT, PT, VI, NC. OUS: Australia, New Zealand, South Africa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.