Recall Z-1728-2022 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Exactech, Inc.

GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH GXL 0 DEGREE LINER 28MM SZ E, REF 130-28-25 b. ACUMATCH GXL 0 DEGREE LINER 28MM SZ F, REF 130-28-26 c. ACUMATCH GXL 0 DEGREE LINER 28MM SZ G, REF 130-28-27 d. ACUMATCH GXL 0 DEGREE LINER 28MM SZ H, REF 130-28-28 e. ACUMATCH GXL 0 DEGREE LINER 28MM SZ J, REF 130-28-29 f. ACUMATCH GXL 0 DEGREE LINER 32MM S

DateAug 11, 2022
StatusOpen, Classified
ReasonSpecific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.
Root causeProcess control
ActionThe firm sent an "URGENT MEDICAL DEVICE RECALL" DEAR HEALTHCARE PROFESSIONAL COMMUNICATION on 08/11/2022 by email. The purpose of the current letter is to update surgeons regarding the implications of the packaging non-conformity on performance of the Connexion GXL and other conventional UHMWPE acetabular liners. In addition, this letter provides updates to patient management and follow-up recommendations since issuance of the July 2021 DHCP letter. In addition to providing surgeons with lists of all their affected GXL and/or conventional polyethylene patients since 2004, Exactech is providi
Quantity11143 DEVICES
DistributionWorldwide distribution: US (nationwide) and OUS (foreign) countries of: ARGENTINA, BRAZIL, CHILE, CHINA, COLOMBIA, ECUADOR, GREECE, GUATEMALA, INDIA, LEBANON PORTUGAL,PUERTO RICO,SPAIN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051556ACUMATCH A-SERIES CONNEXION GXL ENHANCED CROSSLINKED POLYTHYLENE
Exactech, Inc., Sep 6, 2005