ClearanceBrief
Recall Z-1326-2022 ยท KPR System, X-Ray, Stationary

Philips Medical Systems DMC GmbH

DigitalDiagnost 4 High Performance. radiography and fluoroscopy system

DateJul 12, 2022 (initiated May 19, 2021)
StatusOpen, Classified
ReasonFollowing a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy. (Updated 1/30/23) Expansion of the root cause to
Root causeRadiation Control for Health and Safety Act
ActionPhilips sent an Electronic Product Radiation Defect notification letter dated May 25, 2022 to affected customers. Philips will schedule an appointment with customers to install the software update. Philips will, without charge bring the product into compliance. Customers were instructed to complete and return the attached response form to Philips. For further information or support concerning this issue, please contact Philips Customer Care Solutions Center: 1-800-722-9377 and reference FCO71200223 for impacted DigitalDiagnost systems or FCO70600109 for impacted ProxiDiagnost systems.
Quantity140 systems in total, (Updated 1/30/2023).
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K210692DigitalDiagnost
Philips Medical Systems Dmc GmbH, Apr 2, 2021