ClearanceBrief
Recall Z-1691-2022 ยท OEZ Cardiovascular Procedure Kit

American Contract Systems, Inc.

Various Pacemaker Packs, Cardio Thoracic Packs

DateJun 16, 2022
StatusOpen, Classified
ReasonProduct was exposed to multiple sterilization cycles without validation for multiple exposures.
Root causeProcess change control
ActionThe recalling firm began notifying customers of the recall beginning June 13, 2022. Customer notification was initially made via email and an "Urgent Medical Device Recall" letter dated June 16, 2022. Customers are asked to review inventory and segregate affected product, which should be identified with the Bag Serial Number. Affected product should be destroyed, with evidence of destruction provided back to the recalling firm. A Field Action Response Form should be completed and returned to angela.bunn@owens-minor.com as soon as possible, even if the customer has no inventory. If you hav
Quantity24 units
DistributionDistribution within the United States to health systems and distributors in AZ, FL, IA, IL, MA, MN, MO, OH, NE, PA, RI, SD, and TX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.