Recall Z-1525-2022 ยท CBK Ventilator, Continuous, Facility Use

Hamilton Medical AG

HAMILTON-C6 Intensive Care Ventilator, REF: 160021

DateJun 27, 2022
StatusOpen, Classified
ReasonVentilator status indicator board can become loose, which could lead to water ingress (disinfectants) that may lead to technical fault alarms. Multiple technical faults in a short time may force ventilators into Safety Ventilation (blower runs constantly) or Ambient Sate (inspiratory channel/expiratory valves opened; patient breaths room air unassisted) with Panel connection lost message displayed
Root causeDevice Design
ActionOn 6/27/22 Field Safety Notices were emailed to health care facilities, distributors, and service engineers who were informed to take the following actions: 1) In case of a loose status indicator board observed by the user, the front housing containing the status indicator board shall be replaced. This replacement prevents the water ingress that may lead to ambient state. 2) Check your affected ventilator and inform the distributor immediately if you detect the loosening of the status indicator board. 3) Your local distributor or service engineer will contact you shortly. 4) Loose status ind
Quantity358
DistributionUS nationwide distribution including Puerto Rico.