ClearanceBrief
Recall Z-1485-2022 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Cobalt Implantable Cardioverter Defibrillators: a. Cobalt XT VR ICD, Model Numbers: DVPA2D1, DVPA2D4; b. Cobalt XT DR ICD, Model Numbers: DDPA2D1, DDPA2D4

DateAug 9, 2022 (initiated Jun 22, 2022)
StatusOpen, Classified
ReasonThere is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
Root causeDevice Design
ActionBeginning 22-JUN-2022 communications to all affected consignees began. In the United States notifications are sent via UPS 2-day delivery. For outside U.S. countries, notifications are handled via a regionally approved method. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification. Consignees are asked to quarantine the subset of unused identified
Quantity6884 units
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.