Recall Z-1479-2022 ยท OOI Real Time Nucleic Acid Amplification System

Meridian Bioscience Inc

Revogene, Catalog no. 610210. IVD test instrument

DateJun 22, 2022
StatusOpen, Classified
ReasonWhen a run is aborted, the cooling period protection does not occur and the user is able to open the lid prior to the completion of the cooling period, exposing the user to heated instrument components. Incidental contact with heated instrument components may result in an injury such as minor burns.
Root causeDevice Design
ActionOn June 22, 2022, the firm notified customers via email. Customers were instructed that if a run is aborted and an error code presented, they SHOULD NOT open the instrument. DO NOT touch the retention ring or the inside of the lid as they may be hot. The firm recommends that customers wait approximately 10 minutes to allow the instrument to cool prior to opening the lid and removing the test devices. The firm has developed a firmware solution to prevent the lid from being opened after an aborted run prior to the completion of the cooling period. The Meridian Bioscience Technical Services tea
Quantity459 (394 US, 65 OUS)
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Chile, Italy, Qatar, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170558revogene
Genepoc, Inc., May 25, 2017