Recall Z-1478-2022 ยท HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)

Stryker Instruments

Stryker 2.3mm Tapered Router, Catalog no. 5820-071-023 instrument used in craniotomy procedures.

DateJun 24, 2022
StatusOpen, Classified
ReasonThere is a potential for the core diameter to be undersized, which may lead to the router breaking. A break can lead to unintended metal fragments in the surgical site, resulting in the potential for surgical intervention or damage to critical neurological or vascular structures.
Root causeProcess change control
ActionOn June 24, 2022, the firm notified customers via Urgent Medical Device Recall Letters. Customers were instructed to locate and quarantine any affected products found in their inventory. Upon receiving the customer's completed response form, the firm will provide a shipping label to return any recalled product. Upon receipt of the recalled product, Stryker will contact the customer to arrange for replacement product.
Quantity543
DistributionWorldwide distribution - US Nationwide distribution in the states of Arizona, California, Florida, Georgia, Iowa, Michigan, Missouri, New Mexico, North Carolina, Ohio, Texas, Virginia and the countries of Canada, Netherlands, and Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143320Stryker Elite Attachments, Stryker Heavy Duty (HD) Attachments
Stryker Corp., Apr 23, 2015