Recall Z-1508-2022 ยท CGA Glucose Oxidase, Glucose
Alere San Diego, Inc.
The Cholestech LDX System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL cholesterol, triglycerides and glucose. Samples used for testing can be whole blood from a fingerstick (collected in a lithium heparin-coated capillary tube) or venipuncture. The sample is applied to an Cholestech LDX cassette. Description/Part: Cholestech LDX Lipid Profile GLU Cas
| Date | Jun 24, 2022 |
| Status | Open, Classified |
| Reason | Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL). |
| Root cause | Device Design |
| Action | On 6/24/22, correction notices were mailed and emailed to customers who were asked to take the following actions: 1) Test results from patients with bilirubin levels above 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated) should be verified using another test method. 2) Complete and return the Customer Reply Form. 3) If you have forwarded the products listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter. 4) Retain this letter for your laboratory records. Customers with questions are encouraged to contact Technical Servic |
| Quantity | 436,990 |
| Distribution | US: IL, NV, FL, LA, CA, MA, MN, NC, AK, NE, TX, PA, RI, SC, TN, OK, KY, GA, NY, MO, CO, AZ, OH, HI, VA, KS, IN, AL. OUS: GB, ZA, NO, NL, FI, IE, IQ, IT, CH, DE, ES, HU, SI, NA, PT, IS, NG, PL, CY, TN, BH, ID, MY, AU, SG, JP, IN, BD, KR, HK, BR, CO, AR, CA |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K120615 | ALERE CHOLESTECH LDX ANALYZER, ALERE CHOLESTECH LDX LIPID PROFILE - GLU CASSETTE Alere San Diego Inc. Dba Biosite,Innovacon,Hemosen, May 15, 2013 |