Recall Z-1606-2022 ยท JHI Visual, Pregnancy Hcg, Prescription Use

Mckesson Medical-Surgical Inc. Corporate Office

a. ABBOTT RAPID DX N AMERICA LLC Alere Rapid Test Kit Alere Fertility Test hCG Pregnancy Test Urine Sample 30 Tests, Model Number: 92210. b. ABBOTT RAPID DX N AMERICA LLC Alere Rapid Test Kit Alere hCG Dipstick Fertility Test hCG Pregnancy Test Urine Sample 50 Tests Model Number: 92211.

DateAug 26, 2022 (initiated May 25, 2022)
StatusOpen, Classified
ReasonFacility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
Root causeEnvironmental control
ActionMckesson Letter Notifcations issued via USP by or by email if available by the following third-party recall company: Sedgwick Brand Protection (formerly Stericycle) Sedgwick, Memphis, TN 38120 sedgwick@sedgwick.com. from May 10, 2022 continuing through July 21, 2022. Letter states: If you have further distributed this product, please notify your patients at once of this product recall. Your notification to your patients may be enhanced by including a copy of this recall notification letter. This recall should be carried out to the user level. Your assistance is appreciated and necessary to pr
Quantitya. 5 kits b. 12 kits
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062361INNOVACON HCG UKTRA TEST DEVICE
Innovacon, Inc., Oct 23, 2006
K993203ACON ONE STEP PREGNANCY TEST STRIP (URINE)
ACON Laboratories, Inc., Dec 1, 1999