Recall Z-1299-2022 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

RayStation 11B

DateJun 10, 2022
StatusTerminated
ReasonThis notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.
Root causeUnder Investigation by firm
ActionThe firm notified customers of the recall via email on June 10, 2022. The notice was titled "Field Safety Notice, Medical Device Correction #99834." Customers were given precautions and instructed on a workaround. The issue will be resolved in the next version of RayStation, scheduled for market release in June 2022, though this may be subject to market clearance.
Quantity26
DistributionUS Nationwide distribution in the states of NJ and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K220141RayStation 11B
RaySearch Laboratories AB (PUBL), Apr 15, 2022