ClearanceBrief
Recall Z-1426-2022 ยท DQK Computer, Diagnostic, Programmable

Siemens Medical Solutions USA, Inc

Sensis Vibe Combo with software version VD12A. A diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology Model: 11007642

DateJun 15, 2022
StatusOpen, Classified
ReasonSensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication during the first patient examination of the day or after a longer period of inactivity to the Sensis Vibe system resulting in no vital signs being available.
Root causeDevice Design
ActionSiemens issued a Customer Safety Advisory Notice (AX016/22/S) dated 6/15/22 distributed to customers. Letter states reason for recall, health risk and action to take. The BIOS configuration of the ComboBox in the affected systems will be updated to correct the issue. This corrective action will be distributed via Update Instruction AX015/22/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowle
Quantity13 units
DistributionUS Nationwide Distribution AZ FL KY MI OH PR TX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150493Sensis
Siemens Medi Cal Solutions, Inc., Jun 30, 2015