Sensis with software version VD12A. A diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology Model:10764561
| Date | Jun 15, 2022 |
| Status | Open, Classified |
| Reason | Sensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication during the first patient examination of the day or after a longer period of inactivity to the Sensis Vibe system resulting in no vital signs being available. |
| Root cause | Device Design |
| Action | Siemens issued a Customer Safety Advisory Notice (AX016/22/S) dated 6/15/22 distributed to customers. Letter states reason for recall, health risk and action to take. The BIOS configuration of the ComboBox in the affected systems will be updated to correct the issue. This corrective action will be distributed via Update Instruction AX015/22/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowle |
| Quantity | 2 units |
| Distribution | US Nationwide Distribution AZ FL KY MI OH PR TX |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K150493 | Sensis Siemens Medi Cal Solutions, Inc., Jun 30, 2015 |