ClearanceBrief
Recall Z-1456-2022 ยท PDJ Hemoglobin A1c Test System

Abbott Laboratories

REF: 08P4377, Alinity c Hemoglobin Reagent Kit, IVD, CE, R1 5 x 46.0 mL, R2 5 x 18.3 mL, A1cDL 5 x 67.6 mL

DateJun 9, 2022
StatusOpen, Classified
ReasonThere is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using the ARCHITECT and Alinity c Hemoglobin A1c assay.
Root causeUnder Investigation by firm
ActionOn June 9, 2022, Abbott issued a "Urgent Medical Device Correction" to affected consignees via FedEx. 1. Immediately install the updated assay file versions listed below (as applicable for your laboratory), and manually configure the Low-Linearity value. The assay files can be obtained from www.corelaboratory.abbott 2. For the whole blood application, manually configure the THbWB Low-Linearity to 53.8278. For the hemolysate application, manually configure the THbH Low-Linearity to 1250. For detailed information on editing assay parameters, refer to Changing assay configuration settings, Chan
Quantity64 units
DistributionUS: AK AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS MT NC ND NE NJ NV NY OH OK OR PA PR SC SD TN TX UT VA WA WI WV OUS: ALBANIA ALGERIA ANDORRA ANGOLA ARGENTINA ARMENIA AUSTRALIA AUSTRIA AZERBAIJAN BAHAMAS BANGLADESH BELARUS BELGIUM BERMUDA BOSNIA AND HERZEGOVI BRAZIL BRUNEI BULGA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.