Recall Z-1403-2022 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

RayStation 9B SP1. For radiation treatment planning.

DateJun 1, 2022
StatusTerminated
ReasonAn issue with propagation of treatment course information from RayStation with RayTreat to RayCare has been identified. During some workflows, information may not be propagated to RayCare.
Root causeSoftware change control
ActionThe firm contacted all customers via email on June 1, 2022. Customers were provided with a list of workflows where safety related information may not be propagated. Customers were instructed to be aware that treatment course information in RayCare may not always be up to date. Customers should verify critical information in RayStation or RayTreat before treatment course information in RayCare is used for critical decisions. Customers were asked to identify all installed units with the affected software version numbers. The issue will be resolved in mandatory software updates for RayStation,
Quantity1
DistributionWorldwide distribution - US Nationwide distribution in the state of TN and the countries of Belgium and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200569RayStation
RaySearch Laboratories AB (PUBL), Sep 24, 2020