ClearanceBrief
Recall Z-1481-2022 ยท KOA Surgical Instruments, G-U, Manual (And Accessories)

Aesculap Implant Systems LLC

PRESTIGE ATRAUMATIC GRASPER W/PORT DUAL ACTION, ROTATING 5mm DIAM., 36cm LENGTH. Product Code 8360-10

DateApr 1, 2020
StatusCompleted
ReasonEndoscopic graspers may become separated at the weld from handle to the shaft of the device.
Root causeProcess change control
ActionA notification letter was issued April 30, 2020 sent USPS Certified mail . The notification identifies the products as Catalog Item # 8360-10 PRESTIGE ATRA GRASPER DBL-ACT 5MM and 8361-10 PRESTIGE RETR GRASPER DBL-ACT 5MM. Customers were to evaluate inventory with a "tug test" and submit the notification acknowledgment form. Should a device fail the tug test, customers were to file a complaint using Aesculap's Online Customer Complaint Submission Form and return a the product. The products have since been discontinued.
Quantity3392
DistributionUnited States Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Washington DC and Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.