Recall Z-1605-2022 ยท MED Sterilant, Medical Devices

Mckesson Medical-Surgical Inc. Corporate Office

STERIS CORP Revital-Ox Resert Hydrogen Peroxide High-Level Disinfectant Revital-Ox RESERT RTU Liquid 4 Liter Container Max 21 Day Reuse Model Number: 4455AW

DateAug 26, 2022 (initiated May 25, 2022)
StatusOpen, Classified
ReasonFacility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
Root causeEnvironmental control
ActionMckesson Letter Notifcations issued via USP by or by email if available by the following third-party recall company: Sedgwick Brand Protection (formerly Stericycle) Sedgwick, Memphis, TN 38120 sedgwick@sedgwick.com. from May 10, 2022 continuing through July 21, 2022. Letter states: If you have further distributed this product, please notify your patients at once of this product recall. Your notification to your patients may be enhanced by including a copy of this recall notification letter. This recall should be carried out to the user level. Your assistance is appreciated and necessary to pr
Quantity20.25 CS
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080420RESERT XL HLD HIGH LEVEL DISINFECTANT
STERIS Corporation, Sep 2, 2008