Recall Z-1462-2022 ยท LLZ System, Image Processing, Radiological

NordicNeuroLab AS

nordicBrainEX, software versions 2.21 through 2.3.10

DateJun 16, 2022
StatusTerminated
ReasonWhen exporting merged results from the BOLD, DSC, and DCE modules of nordicBrainEx, where both the underlay and overlay originated from a multiframe dataset, the resulting output may have left/right sides flipped.
Root causeSoftware design
ActionThe firm sent a notice regarding a Critical Bug in nordicBrainEx version 2.2.1 to 2.3.10 on 06/16/2022 by email. The notice explained the problem and advised users to use extra caution when viewing the exported results. A software revision will be released to correct the issue.
Quantity243 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133910NORDICBRAINEX
Nordicneurolab, Apr 4, 2014