Recall Z-1597-2022 ยท DOF Lead, Atomic Absorption

Mckesson Medical-Surgical Inc. Corporate Office

MAGELLAN DIAGNOSTICS LeadCare II Blood Lead Test Kit Rapid Test Kit LeadCare II General Chemistry Blood Lead Test Whole Blood Sample 48 Tests Model Number: 70-6762

DateAug 26, 2022 (initiated May 25, 2022)
StatusOpen, Classified
ReasonFacility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
Root causeEnvironmental control
ActionMckesson Letter Notifcations issued via USP by or by email if available by the following third-party recall company: Sedgwick Brand Protection (formerly Stericycle) Sedgwick, Memphis, TN 38120 sedgwick@sedgwick.com. from May 10, 2022 continuing through July 21, 2022. Letter states: If you have further distributed this product, please notify your patients at once of this product recall. Your notification to your patients may be enhanced by including a copy of this recall notification letter. This recall should be carried out to the user level. Your assistance is appreciated and necessary to pr
Quantity81 kits
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052549LEADCARE II BLOOD LEAD TESTING SYSTEM
Esa Biosciences, Inc., Oct 6, 2005