Recall Z-1402-2022 ยท PBZ Image Processing Device For Estimation Of External Blood Loss

Stryker Instruments Div. of Stryker Corporation

Triton Canister System (finished part numbers FG 12009, for 3L canisters, and FG 12019 for 2L canisters) Scanning Label (component part number 11032). A software application used to help estimation of blood loss. Used with blood bearing canisters and accessories.

DateJan 22, 2020
StatusTerminated
ReasonSome Canister Scanning Labels used to calibrate the Triton Canister System have a color variance. Based on a comprehensive analysis, blood loss estimates obtained using these Canister Scanning Labels may result in lower than expected results.
Root causeProcess control
ActionCustomers were notified via email dated 1/22/20. We request the following actions from you: 1. Please confirm receipt of this communication by emailing myself and/or your Gauss representative. 2. Please identify any of these miscolored Canister Scanning Labels in your possession: " Find the the miscolored labels by locating any package of insert + scanning label with a date of manufacture prior to to 2019-12-24. See an example below: 3. Once identified, please: " Quarantine this product and indicate this has been completed by written email response to this communication Your Gauss represe
Quantity16,700 units
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142801Triton Canister System
Gauss Surgical, Inc.,, Mar 12, 2015