Recall Z-1385-2022 ยท BTT Humidifier, Respiratory Gas, (Direct Patient Interface)

Hamilton Medical AG

HAMILTON-H900 Humidifier, Models: 950001, 950004

DateJan 5, 2021
StatusCompleted
ReasonWhen starting a humidifier, intended for respiratory gas conditioning during invasive and noninvasive mechanical ventilation, with an empty chamber, and water is not filled into the chamber, the low water level alarm will be delayed and medical staff are unaware that there is no water in the chamber.
Root causeEmployee error
ActionOn 01/05/21, Field Safety Corrective Action notices were sent to customers who were asked to do the following: Follow the instructions in the HAMILTON-H900 Operator s Manual 624431 section 3.4. Before connecting a new patient to the humidifier; " Verify that the alarms operate correctly. " Ensure that the water supply to the humidifier chamber is functioning properly. Distribute this notification to all end users within your organization. A Medical Representative will be in contact to schedule service with a software update. Customers with additional questions or concerns were encouraged
Quantity203
DistributionUS: WI, FL, CA, NV, PA, TX, PA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163283HAMILTON-H900, HAMILTON-BC8010, HAMILTON-BC4010
Hamilton Medical AG, Sep 20, 2017