Recall Z-1395-2022 ยท GWM Device, Monitoring, Intracranial Pressure

Integra LifeSciences Corp.

CereLink ICP Monitor; Catalog No. 826820, 826820P. Intracranial pressure monitor.

DateJul 22, 2022 (initiated Jun 22, 2022)
StatusOpen, Classified
ReasonIntegra has received complaints associated with ICP readings drifting to -50 mmHg (out-of-range) and may manifest an error message "sensor or extension cable failure!" The firm has found that this is caused by electrical interference from a component of the monitor's circuit board and from the environment.
Root causeDevice Design
ActionA recall notification titled 1st NOTIFICATION - URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION, dated June 22, 2022, was sent by mail. This communicated troubleshooting techniques if an out-of-range reading is encountered when utilizing the Codman CereLink System, while Integra finalized a root cause investigation. A second notification, dated August 23, 2022, for this recall states that Integra is now conducting a removal of Codman CereLink ICP Monitors. Customers are asked to discontinue use of affected devices as soon as possible and remove the devices from service. If an affected device
Quantity1,210 units
DistributionWorldwide Distribution: US nationwide, Australia, New Zealand, Canada, Hong Kong, Pakistan, Georgia, Saudi Arabia, Israel, South Africa, United Arab Emirates, Austria, Belgium, Czech Republic, Denmark, Estonia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Latvia, Lithuania, Nethe
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K210993CereLink ICP Monitor
Integra Lifesciences Production Corporation, Apr 30, 2021