Recall Z-1460-2022 ยท NAY System, Surgical, Computer Controlled Instrument
Intuitive Surgical, Inc.
Da Vinci X/Xi Large Clip Applier, Model 470230 Da Vinci X/Xi Medium-Large Clip Applier, Model 470327 Da Vinci X/Xi Small Clip Applier, Model 47401
| Date | Jul 1, 2022 |
| Status | Open, Classified |
| Reason | Due to the potential for unexpected motion caused by the partial disengagement of the instrument from the system. |
| Root cause | Nonconforming Material/Component |
| Action | On 07/01/2022, Intuitive Surgical sent an "Urgent Medical Device Correction" Letter to customers via FedEx courier informing them there is the potential for unexpected motion while using the da Vinci S/Si and X/Xi Endowrist Clip Applier Instruments. This can happen when one of the da Vinci S/Si and X/Xi Clip Applier grip discs are disengaged from the sterile adapter disc during the engagement routine that occurs prior to insertion of the instrument into the cannula and into the patient. To mitigate this issue, Intuitive Surgical has developed a Clip Applier Manual Self-Test (Appendix A) that |
| Quantity | 90,628 Clip appliers |
| Distribution | Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY. The coun |
| FDA record | Open on fda.gov |