Recall Z-1373-2022 ยท LRO General Surgery Tray

Alcon Research, LLC

Alcon Custom Pak Surgical Procedure Pack: 16998-11 2519351H SALAHUDDIN 2.4 PACK FRESNO SURGICAL HOSP 18392-03 14HDFF CATARACT SWAGEL WOOTTON EYE INSTITUTE 18527-03 14JM57 DOSS CATARACT ALLIANCE EYE SURG PARTNERS 11802-07 14KFAF GANS BASIC PACK NOVAMED EYE SURG CTR 11802-07 14LTRW GANS BASIC PACK NOVAMED EYE SURG CTR 11802-07 14LLME GANS BASIC PACK NOVAMED EYE SURG CTR 14429-07 14HPW3 DR SIEM

DateJun 9, 2022
StatusOpen, Classified
ReasonDue to latex gloves being include in Custom Pak. Kits are supposed to include non-latex gloves.
Root causeProcess control
ActionOn 06/09/2022, the firm sent an "URGENT: VOLUNTARY MEDICAL DEVICE REMOVAL" Letter via overnight mail to customer informing them that specific lots of Alcon Custom Paks were manufactured containing latex gloves and were intended to contain non-latex gloves. Customers are instructed to: 1. Review your inventory to determine if you have any unused affected product within your facility. See Table above for affected Custom Pak. 2. Segregate any unused affected product from your inventory. 3. An Alcon associate will reach out to you directly to arrange for the return and replacement of your affecte
Quantity320 kits
DistributionU.S.: AZ, CA, MO, and NY O.U.S.: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K880961ALCON SURGICAL PROCEDURE PACKS
Alcon Laboratories, Apr 29, 1988