Recall Z-1372-2022 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Siemens Healthcare Diagnostics Inc

epoc BGEM Test Card 25 Test Cards/box. In vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood Product Siemens SMN#/Model #: 10736515

DateJun 1, 2022
StatusOpen, Classified
ReasonDiscrepant (low bias) pO2 results, could be an incorrect diagnosis of hypoxemia and initiation of unnecessary or inappropriate respiratory support.
Root causeUnder Investigation by firm
ActionSiemens has issued an Urgent Medical Device Correction letter POC 22-010.A.US via FedEx beginning on June 1, 2022. An Urgent Field Safety Notice (UFSN) was issued to OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities. Letter states reason for recall, health risk and action to take: Please dispose any inventory of epoc BGEM test card lot 11-22005-50 currently in your possession in accordance with local and state disposal requirements. You may request free of charge replacement product from your local Siemens or di
Quantity263 Boxes
DistributionAL CA CO FL IA ID KS NM OK TX