Recall Z-1365-2022 ยท FPL Incubator, Neonatal Transport

Draeger Medical Systems, Inc.

Draeger Globe-Trotter and Globe-Trotter IC, Model numbers MS20508 and MS20509. neonatal transport system

DateMay 20, 2022
StatusOpen, Classified
ReasonThe firm has initiated this recall to replace component on devices distributed containing a component within the oxygen supply system that is not rated to the working pressure of the system.
Root causeUnder Investigation by firm
ActionOn May 20, 2022, the firm began to send Important Field Safety Notice letters to affected customers. Customers were instructed to avoid using the ventilator portion of the system due to the potential for enriched ambient air rising beyond safe levels until a replacement part can be installed and tested. The firm will send customers a replacement kit and testing which will be completed at no charge to the customer. Draeger Service Representatives will call customers and schedule a service visit to update affected systems.
Quantity38 OUS (no US distribution)
DistributionNo US distribution. Foreign distribution to BY, CL, DE, ES, GB, HU, IN, IT, KW, MX, SA, SV, TW.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141565GLOBE-TROTTER GT5400
Draeger Medical Systems, Inc., Sep 10, 2014