ClearanceBrief
Recall Z-1359-2022 ยท OZP Automated Insulin Dosing Device System, Single Hormonal Control

Medtronic MiniMed

MiniMed 780G Insulin Pump (O.U.S. Version)

DateMay 23, 2022
StatusOpen, Classified
ReasonDue to battery cap deterioration, this may potentially result in an incomplete battery circuit (similar to when battery cap is not attached) and a loss of power/functionality of the insulin infusion pumps.
Root causeDevice Design
ActionOn 05/24/2022, Medtronic began communicating an "URGENT MEDICAL DEVICE CORRECTION" Letter to customers informing them of the battery cap issue where if the metal contact becomes loose or falls off from the battery cap, it can result in an incomplete battery connection, leading to no power source to the insulin pump. Actions Required by customers: Before beginning: Do not remove the battery cap unless you have a new battery available. If you have a spare undamaged battery cap, ensure it is available nearby. During routine battery replacement, check the metal contact on your pump battery cap to
Quantity83,398 pumps
DistributionU.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, and WY O.U.S.: Not provided
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.