Recall Z-1364-2022 ยท FIP Subsystem, Water Purification

Fresenius Medical Care Holdings, Inc.

AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)

DateAug 3, 2021
StatusTerminated
ReasonAquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation
Root causeLabeling Change Control
ActionCustomers were notified via telephone beginning 8/3/21. An FMCRTG, LLC Water Systems Specialist contacted the two consignees via phone and scheduled a time to perform the inspection on the AquaBplus 2000 devices. No formal written communication will be sent since calls and inspections have already performed.
Quantity2 units
DistributionUS Nationwide distribution in the states of NC and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133829AQUABPLUS,AQUABPLUS & AQUABPLUS B2,AQUABPLUS&AQUABPLUS HF,AQUABPLUS& AQUABPLUS B2& AQUABLUS HF
Vivonic GmbH, Apr 22, 2014