Recall Z-1343-2022 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Artis zeego (Model no. 10280959) and Artis Q.zeego (Model no. 10848283), Interventional Fluoroscopic X-Ray System

DateMay 24, 2022
StatusTerminated
ReasonIt may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are blocked and can only be reactivated by a field service engineer.
Root causeComponent design/selection
ActionOn May 24, 2022, the firm mailed Urgent Medical Device Correction letters to affected consignees. Customers were informed that if the system is not used regularly, please ensure that the system is switched on in due time before any use to check the status. Please ensure that patient treatment can be continued in other ways if there is any possible danger for the safety of the patient. The firm will replace the Bios battery and the service interval will be shortened to two years. The firm will contact customers to arrange a date to perform the correction. Customers may call the service organi
Quantity233 worldwide; 32 US
DistributionWorldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CO, FL, IL, LA, ME, MN, MO, NC, OH, PA, SC, TN, TX, WI and the countries of Australia, Austria, Azerbaijan, Belgium, Canada, China, Finland, France, Germany, Hong Kong, India, Italy, Japan, Lebanon, Lithuania, Luxemb