Recall Z-1309-2022 ยท FQO Table, Operating-Room, Ac-Powered

Skytron, Div. The KMW Group, Inc

Product Name: GCX Channel Mounting Accessory Model/Catalog Number: H8-300-40-A

DateJun 2, 2022
StatusOpen, Classified
ReasonUnder certain circumstances, the mounting screws may loosen or pull out from the attachment point on the Ergon 3 boom when the GCX Channel Mounting Accessory was used with the VST clamp as a monitor mount on the boom.
Root causeComponent design/selection
ActionOn June 2, 2022, the firm sent out Urgent Medical Device Correction letters to affected customers. Customers were instructed to identify their GCX Channel Mounting Accessories used with Ergon 3 booms. Customers should inspect their mounting accessories. If the user sees that the channel is tilted on the VST clamp, this could indicate that the screws or tension knob have loosened. Remove the device from service if you find evidence that it may fail. Skytron authorized representatives will contact customers to schedule a service visit. Please contact quality at quality@skytron.com with any que
Quantity731
DistributionWorldwide distribution - US Nationwide and the countries of Canada and Japan.
FDA recordOpen on fda.gov