Recall Z-1340-2022 ยท MMI Immunoassay Method, Troponin Subunit

Siemens Healthcare Diagnostics, Inc.

Dimension LTNI -In vitro diagnostic test intended to quantitively measure cardiac troponin-I levels in human serum and heparinized plasma to aid in the diagnosis of myocardial infarction. Siemens Material Number (SMN)/REF (Catalog Number): 10444896/RF521

DateJul 6, 2022 (initiated May 18, 2022)
StatusTerminated
ReasonPositive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 g/L], may lead to inappropriate intervention for myocardial infarction
Root causeUnder Investigation by firm
Action****Recall Expanded 9-21-22*** US Customers were sent Urgent Medical Device Correction (DC-22-2.B.US) on 09/21/2022 and an Urgent Field Safety Notice UFSN DC-22-02.B.OUS) was issued to all OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities beginning on 09/26/2022. Letter informs them to use serum patient samples only when utilizing lot BA3088. Laboratories unable to use serum patient samples for testing are to request a no charge replacement for their impacted reagent. Siemens issued Urgent Medical Device Correc
Quantity76 units
DistributionWorldwide distribution - US Nationwide distribution in the states of AL, AR, CA, IL MN, MS, NC, ND, NH, OK, SD, TN, TX, WA and the countries of Argentina, Brazil, Canada, France, Germany, India, Mexico, Poland, Portugal, Slovakia, Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.