Recall Z-1511-2022 ยท IZZ Table, Radiographic, Non-Tilting, Powered

GE Healthcare, LLC

Service manuals for the CT, PET, or NM table as follows: GT1700/GT1700V/VT1700, GT2000/VT2000, GT2000x/VT2000x, High Capacity Table, Lite Table, Kunlun Table, Global PET-CT Table, NP Table, P2005AA Table, LCT Table, CTE TWIN TABLE Assy. All devices are patient tables for PET/CT systems.

DateAug 10, 2022 (initiated Jun 2, 2022)
StatusOpen, Classified
ReasonAs a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.
Root causeError in labeling
ActionThe recalling firm issued letters dated 6/2/2022 via traceable means on 6/2/2022 to the following titles within the direct accounts: Director of Clinical / Radiology, Risk Manager / Hospital Administrator, and Director of Biomedical Engineering. The letter explained the safety issue that the table could drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place, resulting in injury to service personnel. The letter reported there is no defect in the table itself and the issue is related to unclear guidance
Quantity24,843 tables
DistributionUS Nationwide distribution including PR. There was also government/military distribution. The countries of Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Plurinational State of Bolivia, Bosnia and Herzegovina
FDA recordOpen on fda.gov