Recall Z-1339-2022 ยท OWB Interventional Fluoroscopic X-Ray System
Siemens Medical Solutions USA, Inc
syngo Application software VE20 (Material Number 10848815) installed on the Artis pheno and Artis icono systems as follows: (1) Artis pheno Model Number:10849000 (2) Artis icono biplane Model Number: 11327600 (3) Artis icono floor Model Number: 11327700
| Date | Feb 25, 2022 |
| Status | Open, Classified |
| Reason | After CT image data from Toshiba is loaded, image mirroring can occur along the horizontal and vertical image axes. If this error occurs, the patient orientation/position may be misinterpreted and result in inappropriate treatment, even if the incorrect visualization is obvious. |
| Root cause | Software design |
| Action | Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice to customers via AX007/22/S dated 2/25/22. Letter states reason for recall, health risk and action to take: We do not consider it necessary to re-examine any patients previously treated using the syngo application software. Siemens will correct the software error via Update Instruction AX006/22/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following th |
| Quantity | 127 units |
| Distribution | US Nationwide. |