Recall Z-1351-2022 ยท LPC Device, Angioplasty, Laser, Coronary
Spectranetics Corporation
Philips Laser System, REF LAS-100, Intermittent Operation, Duty cycle dependent upon device in use 100 - 240 VAC, 50/60HZ, 16 AM
| Date | Jun 1, 2022 |
| Status | Open, Classified |
| Reason | The LAS-100 Laser system may detect an inoperable hardware component during power up, which results in an error code and the system not being operable until code is cleared. |
| Root cause | Device Design |
| Action | On June 08, 2022 Spectranetics issued a "Urgent Medical Device Correction" notification to affected consignees via USPS. The recall notification asked consignees to take the following actions: 1. Philips internal testing confirms that 1-3 reboots of the system will temporarily resolve the issue. Philips also recommends the continued use of the Philips Laser System, and to follow the Operator s Manual, as there is no expected harm to the user or patient. Philips recommends notifying all Philips Laser System users within your facility of this communication and retain a copy of this letter for |
| Quantity | 105 units |
| Distribution | US: AR AZ CA CO CT FL GA HI IA IL IN KS LA MA MD ME MI MN MO NC NJ NY OK PA TN TX VA OUS: Saudi Arabia, Denmark, France, Greece, India, Italy, Netherlands, Serbia, Spain, Sweden, United Kingdom |
| FDA record | Open on fda.gov |