Recall Z-1440-2022 ยท FMG Stopcock, I.V. Set

Becton Dickinson & Company

Nexiva with Single BD Connecta Stopcock Version (Pink) Catalog No. 383687 (OUS)

DateMay 9, 2022
StatusOpen, Classified
ReasonMay have the potential for leakage at the housing component of the stopcock, may result in delay or interruption in treatment, exposure to infusate and biohazardous material, under dosing/under infusion, contamination and/or air ingress.
Root causeUnder Investigation by firm
ActionBD notified consignees Medical Device Field Correction Notice on 5/9/22 via mail/email. Firm issued Urgent Medical Device Field Correction Letter extended to additional lots on July 13, 2022 via email. Letter states reason for recall, health risk and action to take: Please Take the Following Actions: 1. Inspect your inventory, locate and quarantine any units of the impacted lot numbers and destroy all affected product. 2. For Clinical Users: For devices in situ, check to see if the device is within scope of this Field Safety Notice and replace immediately. If you are unable to determine if t
Quantity40,050
DistributionWorldwide distribution - US Nationwide distribution in the states of AK, AR, AZ, CA, CO, FL, GA, IL, IN, KS, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OR, PA, SD, TN, TX, UT, VA, WA, WV and the countries of Belgium, China, India, Indonesia, Japan, Korea, Nepal, Philippines, Singapore, Sri Lank
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K974083CONNECTA PLUS 1 AND CONNECTA PLUS 3 2-WAY AND 3-WAY STOPCOCKS
Ohmeda Medical, Jan 13, 1998