Recall Z-1306-2022 ยท ILQ System, Communication, Powered

Baxter Healthcare Corporation

NaviCare Nurse Call/Voalte Nurse Call

DateMay 31, 2022
StatusOpen, Classified
ReasonAn issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority.
Root causeSoftware Design Change
ActionThe firm disseminated an Urgent Medical Device Correction Notice beginning on 05/31/2022 by letter. The letter described the problem and requested the following actions be taken by the customer: "Identify if your Nurse Call System utilizes one of the impacted wireless integrations identified. If so, please share the information with potential users in your organization. Hillrom recommends reviewing the wireless system settings with your wireless provider to determine if there are configurable options for call acceptance only. Complete the enclosed response form and return to hillromCARY021@
Quantity283 installations
Distributionnationwide and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.