Recall Z-1381-2022 ยท PHC Infusion Safety Management Software

Ivenix, Inc.

Ivenix Infusion System (IIS)

DateApr 22, 2022
StatusTerminated
ReasonDownstream occlusion alarm is sometimes immediately followed by a pump problem alarm.
Root causeSoftware design
ActionOn April 22, 2022, the firm notified affected customers of the product issue through a letter titled "Important Safety Update." The firm is working on a software update to correct the issue. It will be available in approximately 60 days. The firm will inform customers when the update is ready to install. As interim corrections, the firm recommends that any pump that is infusing into the downstream access port have an infusion pressure less than 525 mm Hg. Alternatively, the Occlusion Retry Attempts in the Infusate Library Care Profile Pump Setting should be set to 0. Care profile pump settin
Quantity1335 devices
DistributionUS Nationwide distribution in the states of NJ and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183311Ivenix Infusion System (IIS)
Ivenix, Inc., Jun 7, 2019