Recall Z-1295-2022 ยท OEZ Cardiovascular Procedure Kit

ROi CPS LLC

regard Item Number: 800774001, CV00979A - CV Minor Kit, Sterile.

DateJun 28, 2022 (initiated May 20, 2022)
StatusTerminated
ReasonA nonsterile component was packaged in a sterile convenience kit.
Root causeProcess change control
ActionThe recalling firm called the consignee on 5/20/2022 to discuss the recall, including quarantine of affected product, customer fulfilment/available inventory, end user facility notification, recall timelines, and establishing a plan for relabeling by the recalling firm personnel at the distribution center. The recalling firm issued an email to the consignee on 5/20/2022 with an attached recall notice that was previously discussed with the consignee. The email requested the consignee to quarantine the kits and immediately notify the end user facilities who have been shipped the affected kits.
Quantity180 kits
DistributionUS Nationwide distribution in the state of FL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.