ClearanceBrief
Recall Z-1321-2022 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Radiometer Medical ApS

ABL800 Flex Analyzer Model Numbers 393-800 and 393-801

DateMay 18, 2022
StatusOpen, Classified
ReasonThere is a potential for sporadic incidents of positive and negative biases for analyzer systems configured with cNa+, cCa+, and cK+.
Root causeSoftware design
ActionOn May 20, 2022, Radiometer issued a "Urgent: Medical Device Recall" notification to affected consignees via: Email and FedEx 2nd day Mail. Radiometer ask consignees to take the following actions: 1. For the affected micro measuring modes Radiometer kindly requests you to not use results for cNa+. The most efficient way to ensure this is to remove the parameter from the affected modes as per the procedures included within the letter. 2. Please complete the Recall Response Form and return to your Radiometer representative (recall@radiometeramerica.com).
Quantity1489 analyzers
DistributionWorldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Bermuda, Canada, Italy, Norwa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041874ABL800 FLEX
Radiometer Medical Aps, Oct 8, 2004