Recall Z-1265-2022 ยท DXJ Display, Cathode-Ray Tube, Medical

GE Healthcare, LLC

CARESCAPE Central Station (CSCS) V2, Model numbers 2082278-001 and 2082279-001. Update 3/10/2023: The following FRU and display part numbers have been added as affected devices: (1) Part #2082294-001, FRU MP200 POWER SUPPLY ASSEMBLY; (2) Part #2082293-002-R, FRU MP200 CPU ASSEMBLY; (3) Part #2082293-002, FRU MP200 CPU ASSEMBLY; (4) Part #2071523-001, MP200 22 IN DISPLAY WITH TOUCH; (5) Part #20

DateJun 13, 2022 (initiated May 10, 2022)
StatusOpen, Classified
ReasonThe CARESAPE Central Station (CSCS) V2 can shut down due to a potential power supply component failure. This can lead to loss of patient monitoring at the central station.
Root causeProcess control
ActionThe recalling firm issued letters flagged URGENT MEDICAL DEVICE CORRECTION dated 5/10/2022 via traceable means. The letter listed the affected product details and described the safety issue and actions to be taken by the customer/user. It said they can continue to use the CSCS V2 to monitor patients on the CARESCAPE Network. In the event there is a loss of patient monitoring at the CSCS V2, they were to follow the safety instructions, which included: Ensure all patients are adequately monitored using other GE Central Stations, with alternate monitoring devices, or with increased patient sur
Quantity30,993 Central Stations and 2,180 FRUs and displays
DistributionWorldwide distribution - US Nationwide. There was also military/government distribution and the countries of Algeria, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Cameroon, Canada, Chile, China, Colombia, Costa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162012CARESCAPE Central Station V2