Recall Z-1275-2022 ยท FDS Gastroscope And Accessories, Flexible/Rigid

Olympus Corporation of the Americas

EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model: GIF-HQ190

DateApr 22, 2022
StatusOpen, Classified
ReasonPart (A-rubber) intended for a different model of gastrointestinal videoscope may have been used when replacing part on GIF-HQ190 during the last repair at Olympus. An Incorrect A-rubber fits tighter on the GIF-HQ190 which would not meet the intended design specifications
Root causeNonconforming Material/Component
ActionOlympus issued Urgent Medical Device Correction letter via email on 4/22/22. Letter states reason for recall, health risk and action to take: 1. Identify the affected model and serial number in the subject of this letter. 2. An Olympus representative will contact you to arrange for the return of your affected device. Olympus will repair your unit free of charge. 3. Olympus will provide for a service loaner in an effort to not disrupt patient care. Please do not hesitate to contact @ 647-999-3203 or at Cynthia.Ow@Olympus.com for any additional information or support concerning this matter.
Quantity8 units
DistributionUS Nationwide distribution in the states of AL, FL, NC, NY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.