Recall Z-1284-2022 ยท MLM Enzyme Immunoassay, Tracrolimus

Microgenics Corporation

QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL C 10015568-3 QMS TACROLIMUS CAL D 10015569-3 QMS TACROLIMUS CAL E 10015570-3 QMS TACROLIMUS CAL F 10015571-3

DateJan 20, 2022
StatusOpen, Classified
ReasonDue to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay.
Root causeProcess control
ActionOn 01/20/2022, the firm contact via email its international customers regarding that the QMS Tacrolimus Calibrator was recovering between 82 to 94% of its target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay. On 03/10/2022, the firm notified its U.S. customer of the issue. Customers were instructed to: 1)Determine if they are using or have inventory of QMS Tacrolimus Calibrator catalog number 10015573, lot 74175462 or lot 74267412. 2)Discontinue use of affected in-date lots, and transition to new lot(s) where correcti
Quantity615 kits
DistributionU.S.: AL, AZ, CA, DC, FL, IA, ID, IL, KY, LA, MA, NE, NJ, NM, OH, OK, PA, SC, TN, TX, and VA O.U.S.: Bangladesh, Brazil, Canada, Germany, India, Malaysia, Mexico, South Korea, and Thailand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123343THERMO SCIENTIFIC QMS TACROLIMUS ASSAY AND CALIBRATORS
Microgenics Corp., Jul 11, 2013