Recall Z-1246-2022 ยท OWB Interventional Fluoroscopic X-Ray System
Siemens Medical Solutions USA, Inc
Artis icono floor-angiography systems developed for single and biplane diagnostic imaging and interventional procedures SMN: 11327700
| Date | Apr 11, 2022 |
| Status | Open, Classified |
| Reason | if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact with metallic parts of the Carm |
| Root cause | Device Design |
| Action | Siemens Medical Solutions USA, Inc. has initiated a Customer Safety Advisory Notice to all affected customers via AX027/21/S on 4/11/22. Siemens will conduct an inspection of the protective earth via Update Instruction AX026/21/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent |
| Quantity | 13 units |
| Distribution | US Nationwide distribution. |
| FDA record | Open on fda.gov |