ClearanceBrief
Recall Z-1223-2022 ยท GWF Stimulator, Electrical, Evoked Response

Medtronic Xomed, Inc.

NIM Vital Console 4.0 (P/N: NIM4CM01) NIM with Software version 1.3.2

DateJun 2, 2022 (initiated May 3, 2022)
StatusOpen, Classified
ReasonSoftware anomaly was identified.
Root causeSoftware Design Change
ActionThe firm issued a recall notification to consignees on 05/03/2022 via mail. The letter instructs the consignee to follow the IFU regarding the setup and "Lead Off Detected" error message, forward the notification to those who need to be aware within the organization and to any location which the devices have been distributed, and to complete the response form. An additional correction notification was sent on about 09/13/2023 to notify consignees of the update availability (NIM Vital" SW version v1.4.3) to fix this issue.
Quantity18 units
DistributionWorldwide distribution - US Nationwide distribution in the states of TX, CO, NJ and the country of France.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200759NIM Vital, Nerve Integrity Monitor
Medtronic Xomed, Inc., Oct 28, 2020