Recall Z-1248-2022 ยท MEC Disinfector, Medical Devices

Getinge Usa Sales Inc

Getinge CM320-2 CHAMBER WU Series Washer Disinfector- For Use as a multi-chamber washer-disinfector for the washing, disinfecting and drying surgical instruments, etc. Model: CM320 WU

DateApr 13, 2022
StatusOpen, Classified
ReasonVerification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury or slip and fall injury. May lead to a patient procedural delay and/or infection
Root causeProcess control
ActionGetinge issued to U.S. consignees Medical Device Correction Letter dated April 13, 2022 via FedEx delivery. Letter states reason for recall, health risk and action to take: Please examine your inventory immediately to determine if you have any Getinge CM320-series Washer Disinfectors with the REF number/serial numbers listed on page 1. You can continue to use Getinge CM320-series Washer Disinfectors until the corrective action is performed; there are no precautionary actions necessary. 2. Please complete and sign the attached MEDICAL DEVICE CORRECTION - RESPONSE FORM (page 5) to acknowledge t
Quantity4 units
DistributionUS Nationwide distribution in the states of MD, MN, ND, NE, NY, PA, VA.
FDA recordOpen on fda.gov