Recall Z-1225-2022 ยท JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented

MicroPort Orthopedics Inc.

PROFEMUR PLASMA Z CLASSIC STEM SIZE 7 STANDARD SHORT NECK

DateMay 4, 2022
StatusTerminated
ReasonOne manufacturing lot is missing the plasma-spray coating in the proximal region of the stem body.
Root causeProcess change control
ActionThe firm sent out a recall notification to the distributors on 05/04/2022 via email. Phone calls were also made to the distributors. Additionally, the US field communication letter was mailed to the surgeon and facility in regard to the one case of implantation on May 10, 2022. The consignee letter informed the consignee to cease use and distribution of the affected product, check inventory for affected product, complete the response form, return affected product and notify affected consignees. The field surgeon letter and hospital letter informs the consignee to cease use and distribution of
Quantity8 units
DistributionWorldwide distribution - US Nationwide distribution in the states of FL, IN, MO, CA and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123434PROFEMUR Z CLASSIC STEMS
Wrightmedicaltechnologyinc, Feb 5, 2013