Recall Z-1229-2022 ยท LLZ System, Image Processing, Radiological

GE Healthcare, LLC

Centricity Universal Viewer Zero Footprint Client

DateJun 3, 2022 (initiated Apr 28, 2022)
StatusOpen, Classified
ReasonPotential for Distance and Area measurements to display inaccurate measurement values when performed on magnified images, and specific to ZFP, Distance and Area measurements can display inaccurate measurement values when performed on lossy images that are scaled down from their original resolution.
Root causeSoftware design
ActionGE Healthcare notified consignees on about 04/28/2022 via letter titled "URGENT MEDICAL DEVICE CORRECTION." The letter outlined two safety issues and instructed customers to continue to use the device in accordance with the User Manuals, but for issue #1, do not rely on measurements displayed in the viewer. They should manually calibrate the image to create a measurement calibration reference and then perform necessary measurements. For issue #2, do not perform measurements on lossy images. In both instances, customers were requested to complete and return the acknowledgment form to Recall
Quantity488 units
DistributionWorldwide distribution. Nationwide distribution to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. International distribution to Argentina, Aust
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131977CENTRICITY UNIVERSAL VIEWER ZERO FOOTPRINT CLIENT
GE Healthcare, Sep 17, 2013