Recall Z-1242-2022 ยท KRD Device, Vascular, For Promoting Embolization

Cook Incorporated

Tornado Embolization Microcoil, RPN GPN MWCE-18S-5/2-TORNADO G08356 MWCE-18S-6/2-TORNADO G08259 MWCE-18S-3/2-TORNADO-01 G47416 MWCE-18S-4/2-TORNADO-01 G47417 MWCE-18S-3/2-TORNADO-081800 G13102 MWCE-18S-3/2-TORNADO G08261 MWCE-18S-4/2-TORNADO-081800 G13103 MWCE-18S-4/2-TORNADO G08357 MWCE-18S-5/2-TORNADO-01 G47418 MWCE-18S-6/2-TORNADO-01 G47419 MWCE-18S-6/

DateMay 5, 2022
StatusOpen, Classified
ReasonThe loading cartridges included with these devices may contain a small, unintended, stainless-steel cannula.
Root causeUnder Investigation by firm
ActionOn May 5, 2022, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were advised that the presence of a small stainless steel cannula inside the loading cartridge may go undetected by the user. Customers were instructed to examine their inventory and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. Customers should return the product to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Please share this information with appropriate
Quantity106,033
DistributionDomestic distribution nationwide. Foreign distribution worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K153778Nester Embolization Coils, Tornado Embolization Coils
Cook Incorporated, May 19, 2016