Recall Z-1267-2022 ยท NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Cobalt Models loaded with CareLink SmartSync Device Manager application software D00U005: a) Cobalt XT VR: DVPA2D1, DVPA2D4; b) Cobalt VR: DVPB3D1, DVPB3D4; c) Cobalt XT DR: DDPA2D1, DDPA2D4; d) Cobalt DR: DDPB3D1, DDPB3D4; e) Cobalt XT HF: DTPA2D4, DTPA2D1 f) Cobalt HF: DTPB2D4, DTPB2D1; g) Cobalt XT HF Quad: DTPA2QQ, DTPA2Q1; h) Cobalt HF Quad: DTPB2QQ, DTPB2Q1
| Date | Apr 12, 2022 |
| Status | Open, Classified |
| Reason | Telemetry error that may occur with Medtronic Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds). |
| Root cause | Software design |
| Action | A Recall notification was sent to the consignees on 04/12/2022 via the United Parcel Service. The letter instructs the consignee to continue normal follow-up frequency per local clinic protocol and for the consignee to submit the response form. |
| Quantity | 51922 units |
| Distribution | Distributed nationwide to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and internationally to Andorra, Australia, Austria, Belgium, Brunei Daru |
| FDA record | Open on fda.gov |