ClearanceBrief
Recall Z-1130-2022 · ITI Wheelchair, Powered

Invacare Corporation

Invacare TDX SP2, Model Nos. TDXSP2 TDXSP2-MCG TDXSP2V

DateMar 30, 2022
StatusOpen, Classified
ReasonThe battery wiring harness on affected wheelchairs may become disconnected during shipment from the factory or during wheelchair use traversing large obstacles or after a jolt or impact to the wheelchair. The disconnection of the battery harness results in loss of power to the wheelchair.
Root causeDevice Design
ActionA preliminary notification was sent to providers on March 30, 2022, introducing the product issue and providing instructions regarding how to handle impacted units before field correction kits are available. A notification letter dated April 13, 2022, was sent to impacted consignees via traceable mail. This mailing includes a Provider Letter which will introduces the field correction requirements and provides instructions for acknowledging awareness and understanding of the requested field correction activities. Providers should inform affected customers to inform them that their wheelchair
Quantity133 devices
DistributionWorldwide distribution - US Nationwide distribution in the states of AK, AR, CA, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA, and WI. The country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170507Invacare® TDX® SP2 Power Wheelchair; TDXSP2 TDX SP2 with LiNX Electronics, TDXSP2-CG TDX SP2 Base for Single Act. System
Invacare Corporation, Jul 12, 2017